10
MA601990_20 [Original Instructions] 10/04/2010
Water Connections
1. Connect one end of the water supply hose to the water supply inlet on the grinder polisher base. A
water solenoid (provided with the Vector Power Head) is needed to connect the water supply hose.
The water solenoid is marked as “1” and the outlet is marked as “2”.
2. Connect the other end of the water supply hose to an external water source equipped with a separate
shut-off valve. A ½-inch NPT connector is required.
Figure 12 Water Solenoid
Air Connections
The Vector Power Head is shipped with a filter capable of filtering a range of 55 psi (3.8 bar) - 85 psi
(5.8 bar). The air supply must be clean, dry, and oil-free.
1. Connect the stem fitting (P/N 824122) of the supplied air filter line adaptor assembly
(see Figure 14) to the air intake on the back of the Vector Power Head (see Figure 13).
2. Connect the supply air line to the filter (P/N R10486) on the other end of the air filter line adaptor
assembly. A ½-inch NPT fitting is required.
Figure 13 Vector Power Head Air, Water, and Electrical Connections
Power Cord
Electrical Connector for the
Water Solenoid
Electrical Connector to
Grinder Polisher Base
Air Intake
Electrical Connection for
Water Solenoid
Connection to Grinder
Polisher Base