2 electrodes
G ¾'' connection
5-pin M12 male connector
Article number 570452 570453
Measurement range
0.05...20 µS.cm
–1
1...200 µS.cm
–1
Linearity
1)
(relative)
±0.5...5 %
Cell constant
2)
0.01 cm
–1
0.1 cm
–1
Fluid temperature
–20...+150 °C
Fluid pressure
PN16 at –20...+120 °C
PN10 at +120...+150 °C
Temperature sensor
Pt1000
Electrical connection
5-pin M12 male connector
1) Uncertainty of ±5 % arises when using only one single cell constant for
the full range. ±0.5 % measurement deviation can be achieved when
calibration is performed in a conductivity range close to that of the used
solution.
2) Nominal cell constant.
Every product is measured according to a Bürkert standard procedure
and the individual cell constant measured is reported in the calibration
report, delivered with the product. The cell constant can be influenced by
the assembly situation.
2 electrodes
NPT ¾'' connection
5-pin M12 male connector
Article number 570454 570455
Measurement range
0.05...20 µS.cm
–1
1...200 µS.cm
–1
Linearity
1)
(relative)
±0.5...5 %
Cell constant
2)
0.01 cm
–1
0.1 cm
–1
Fluid temperature
–20...+150 °C
Fluid pressure
PN16 at –20...+120 °C
PN10 at +120...+150 °C
Temperature sensor
Pt1000
Electrical connection
5-pin M12 male connector
1) Uncertainty of ±5 % arises when using only one single cell constant for
the full range. ±0.5 % measurement deviation can be achieved when
calibration is performed in a conductivity range close to that of the used
solution.
2) Nominal cell constant.
Every product is measured according to a Bürkert standard procedure
and the individual cell constant measured is reported in the calibration
report, delivered with the product. The cell constant can be influenced by
the assembly situation.