2 electrodes
1½'' clamp connection
Cable
Article
number
564898 562261 564899
Measurement
range
0.05...20 µS. cm
–1
1...200 µS. cm
–1
5...5000 µS.cm
–1
Linearity
1)
(relative)
±0.5...5 %
Cell
constant
2)
0.01 cm
–1
0.1 cm
–1
1 cm
–1
Fluid
temperature
max. +120 °C
Fluid
pressure
max. 7 bar (100 psi)
Temperature
sensor
Pt1000
Electrical
connection
Cable, length 3 m, instrument side with open wire.
1) Uncertainty of ±5 % arises when using only one single cell constant for the
full range. ±0.5 % measurement deviation can be achieved when calibration
is performed in a conductivity range close to that of the used solution.
2) Nominal cell constant.
Every product is measured according to a Bürkert standard procedure
and the individual cell constant measured is reported in the calibration
report, delivered with the product. The cell constant can be influenced by
the assembly situation.
4 electrodes
1½'' clamp connection, short or
long insertion depth
VarioPin male connector
Article number 562420 (short), 564064 (long)
Measurement range
0.1...500000 µS.cm
–1
Linearity
1)
(relative)
±0.5...5 %
Cell constant
2)
0.147 cm
–1
Fluid temperature
–20...+135 °C
Fluid pressure
max. 6 bar (87 psi)
Temperature sensor
Pt1000
Electrical connection
VarioPin (VP 6.0) male connector
1) Uncertainty of ±5 % arises when using only one single cell constant for
the full range. ±0.5 % measurement deviation can be achieved when
calibration is performed in a conductivity range close to that of the used
solution.
2) Nominal cell constant.
Every product is measured according to a Bürkert standard procedure
and the individual cell constant measured is reported in the calibration
report, delivered with the product. The cell constant can be influenced by
the assembly situation.