Installation
Please ensure you have familiarised yourself with the safety statements before installing this equipment.
For optimum performance, the instrument should be installed
according to the following conditions:
The environment must be clean and free of dust and airborne
contaminants.
The instrument should not be stored beneath overhanging
cupboards. Allow a minimum of 50cm of clear space above
the chimney.
The instrument must be placed on a sturdy worktop. The
BWB model range requires approximately 50cm deep by
60cm wide by 100cm high of bench space.
Avoid sites that expose the instrument to direct sunlight or
draughts.
To meet the specification, the ambient temperature must be
within the range +10°C to +35°C and a maximum relative
humidity of 85%, non-condensing.
An AC supply of 100V to 250V, at 50 or 60Hz, is required for the
BWB model range. The power supply automatically detects the
mains voltage and provides the correct power to the various
components of the instrument. The maximum current drawn is
2 amps. Only the BWB supplied DC power adaptor should be
used. Any other DC supplies may invalidate the warranty or void
the CE testing and certification. The instrument can be run from
a 12vDC battery (such as those found in a car) with a suitable
adapter cable available from BWB. Please contact us should this
be required.
A gas fuel supply of Propane, Butane, or Propane/Butane
mixture (as in LPG) is used on the BWB model range. It should
be regulated at the source to no more than 20Bar with a
minimum flow rate of 0.4 litres/minute. The use of industrial
quality gas is not recommended as impurities can enter, leaving
deposits of dirt and oil, which will render the instrument
inoperable.
Attach the supplied high-pressure hose* between the gas
source and the gas regulator on the rear panel using the clamp
provided to ensure a good seal.
*Refer to the safety statements.
It is recommended that a leakage check be performed using a
soap solution or a BWB leak detection solution the first time the
instrument gas supply is turned on to ensure the integrity of the
connections.
13