Calibration
It is recommended that calibrations be performed frequently for
optimum accuracy, particularly if the instrument has been unused
for a lengthy period of time. Factors such as change in room
temperature, humidity and ambient air pressure can affect
calibration and impair the accuracy of results.
Do not attempt to calibrate the instrument until the warm-up
period has elapsed.
The display will show the remaining warm-up period based on a
nominal 45 minute interval.
The flame inspection port must not be opened during calibration or
measurement.
A calibration must be performed prior to any sample
measurements being undertaken.
The display will prompt the operator during many of the
procedures. At any time if an error is made, press the back/STOP
key to return to the previous step.
Single point calibration
1>Read
2 Auto Read
3 Calibrations
4 Set Up
1 View
2>Calib
3 Correct
4 N/A
1 Na None 5 Ba None
2>K None
3 Li. ppm
4 Ca: ppm 8 Multi
Select Calibrations from
the main menu
Select Calib
This screen displays the ions that have calibrations already
assigned. A dot (.) after the element shows the stored calibration
is single point; a colon (:) after the element indicates a multi-point
calibration (in this example Li is shown as a single point, Ca as a
multi-point).
Select the ion to be calibrated (in this example K) by pressing the
relevant numeric key or scroll and press accept. For a single
calibration enter 1 (The value displayed in brackets refers to the
previous calibration data points, if this is still correct, press accept
instead of entering a new number) and press accept.
K Calibration- ppm
Number of
Points (0):
K Calibration- ppm
Number of
Points (0): 1
K Calib- Low Cal
Aspirate Blank
Flushing chamber . . .
--- -
A screen offering the option to skip is displayed, this allows
calibration adjustment (described later) wait 4 seconds. Ensure
you are aspirating the Blank calibration standard. The screen will
update and start the Blank calibration process.
22
accept to Skip…
OR WAIT . . .
to calibrate