Calibration
View Calibration
Once the required number of calibrations has been completed, the
View option in the Calibrations menu will be available. Select
option1>View and press accept
1>View
2 Calib
3 Correct
4 N/A
1 Na None 5 Ba None
2 K None
3>Li. ppm
4 Ca: ppm 8 Multi
View Calibration will display the calibration state of each ion, the type
of calibration – single point (.), multi-point (:) or none and the units
that the calibration was performed in.
Scroll to the appropriate ion and press accept, or use the numerical
keypad to select the relevant number option, in this example 3>Li.
ppm is shown below.
Li 1 Point R2=1.000
Blank: 385 Raw
01: 10.0 ppm
1720 Raw
This will show a screen displaying the calibration details for that ion:
• The number of calibration points
• The raw reading for Blank
• The concentration for calibration point 1
• The raw reading for calibration point 1
For Multi Point calibrations the up and down arrows will step
through each calibration point in turn.
Pressing accept will move to the next ion.
Pressing the back/STOP key returns the display to the main
calibration screen.
The numeric keys can be used to navigate to a particular ion/
channel (e.g. 1= Na, 2= K, 3= Li, 4= Ca, 5= Ba).
28