Calibration
As with the Multi Point calibration, the standards should be made so
each ion will start with its lowest value and increasing up from that. For
example:
Calibration point
Na
K
1
100
10
2
50
50
3
10
100
Incorrect X
Calibration point
Na
K
1
10
10
2
50
50
3
100
100
Correct