Calibration
Calibration Correction
The BWB model range enables the calibration of any or all Ions to be
corrected to eliminate the effects of drift in the Ion readings. This
option is available only after an Ion has been calibrated.
To initiate Calibration Correction from the Calibrations menu, select
option3>Correct and press accept.
1 View
2 Calib
3>Correct
4 N/A
1 Na None 5 Ba None
2 K None
3>Li. ppm
4 Ca: ppm 8 Multi
Scroll to the ion for which calibration is to be corrected and press
accept, or use the numerical keypad to select the relevant number
option.
For Na, K, Li and Ba ions two correction options will be displayed.
Some additional items are available for Ca which we’ll cover shortly.
1>Correct Blank [Li]
2 Correct Calib
Correct Blank enables the user to eliminate the effects of baseline drift.
Correct Calib enables the correction of the highest calibration standard
in a multipoint calibration, by means of one single point calibration. The
instrument will automatically re-calculate the calibration curve for the
selected Ion
To perform a Blank correction, start aspirating the Blank calibration
standard and select option 1 and press accept.
Li Correct- Blank
Aspirate Blank
Flushing Chamber…
--- · 9
To perform a correction of the highest calibration standard, aspirate
the calibrations standard and select option 2 and press accept. When
aspirating the calibration solution the instrument will undergo a
flushing routine prior to storing once a stable signal is achieved. A
timer is displayed; approximate times can vary from 10 to 60 seconds
dependent on stability of conditions.
Li Correct- Pt 1
Asp 10.0 ppm
Flushing Chamber…
--- · 9
29