Calibration
Calibration Adjustment
The calibration adjustment feature allows the user to make adjustments
to a single calibration point within a calibration curve without having to
complete the entire calibration process. This is particularly useful if the
user notices an error on the R2 value of the calibration curve or if the
wrong calibration standard was accidentally aspirated during a
multipoint calibration.
The user should be familiar with the single and multi point calibration
process.
For this example we have reviewed the calibration data for our 5 point
calibration and notice that we calibrated position 4 incorrectly.
We can re-calibrate this single point without having to complete an
entire re-calibration.
Navigate to Calibrations > Calib and select the same parameters as you
had previously (multi or single element). In this example we conducted
a 5 point Potassium (single ion) calibration.
The system is preconfigured with the 5 points so we can press Accept.
K Calibration- ppm
Number of
Points (5):
It is now important to press the Accept button to Skip the first
process of calibration (Blank). This allows us to skip forward to the
next calibration point without having to conduct a Blank
calibration.
K Calibration- Pt 1
accept Preset or
Enter Concentration
(5.0ppm):
The system is pre-configured with the calibration concentration
for point 1 so press Accept and then press Accept again to skip the
process of calibration position 1.
K Calibration- Pt 2
accept Preset or
Enter Concentration
(10.0ppm):
Continue pressing Accept until you reach the calibration point you
wish to Adjust (re-calibrate), in this instance Calibration point 4.
32
accept to Skip…
OR WAIT . . .
to calibrate
accept to Skip…
OR WAIT . . .
to calibrate