Calibration
Note: If during either Blank or Calibration correction the signal value is
outside the acceptable range for the solution being aspirated, a
warning message will be displayed.
Li Correct- Pt 1
Asp 10.0 ppm
Acquiring value …
Wrong Soln! ACCEPT?
Check that the correct solution (Blank or Calibration) and/or the
correct calibration concentration is being aspirated. The correction
can either be accepted by pressing the accept key (although excessive
drift should be suspected and investigated), or aborted by pressing the
back/STOP key.
One-key Calibration Correction
In Read mode the blank or calibration correction can be carried out
with a one-key method, saving time. Aspirate the blank or the highest
calibration standard. Observe the stability indication line and press the
blank or calibrate key respectively when a stable signal is indicated.
The blank or calibration correction will be performed in the same way
as above.
Note: Care must be taken to ensure the blank or calibrate keys are
not pressed accidentally while running a sample as doing so will
invalidate the result.