The setup menu
1>Config 5 Maint
2 Interfaces
3 Ions
4 Test Hardware
1>LCD Contrast: 40
2 Setup Chart
3 AFHS (OFF)
4 Printer (OFF)
The interface menu
Set LCD contrast- cycle through by pressing the accept key.
Setup Chart- used to configure the analogue output for use with a
chart recorder or mA output depending on your model / optional
extras fitted.
AFHS (ON/OFF)- if the Automatic Fluid Handling System (AFHS) is
present this should be set to ON and the Dwell Time should also be
set. (Refer to AFHS Manual).
Printer (ON/OFF)- if the Printer accessory is fitted this should be set
to ON in order to print.
1>Config 5 Maint
2 Interfaces
3 Ions
4 Test Hardware
1>Ion Resolution
2 Next Calib: ppm
1>Na 1dp 5 Ba 1dp
2 K 1dp
3 Li 1dp
4 Ca 1dp 8 All
Ion Resolution:
This mode changes the amount of decimal places that will be shown on
the Ion Readings. The factory default is 1 decimal place. This is optimum
for most applications. Increasing the decimal places will make the
readings appear to be less stable.
1>Ion Resolution
2 Next Calib: ppm
1>ppm 5 %
2 mg/l
3 meq/l
4 mmol/l 8 mg/kg
Next Calib:
Enables the operator to set the units of measure. Any change takes place
on the next calibration. Existing calibrations are not changed and will
remain in their calibrated units.
20