Residual Risk
Low (L) / Medium (M) / High (H)
Flammable / Explosive Substances
Gas (propane, butane, natural
gas)leak leading to explosion
Local exhaust ventilation covering the region where
equipment is to be used.
Upon initial connection of gas tube and propane tank, joints
are to be checked with a “digital gas sniffer”, leak detection
spray or other suitable test.
Joints are to be checked for “tightness” prior every ignition
of the instrument.
Exposure to Hazardous Substances
Some samples / standards may
be hazardous
Follow safe working procedures for the correct handling of
chemicals, samples or standards based on material safety
data sheets.
PPE, Gloves and safety glasses are to be worn in the vicinity
of the Flame Photometer or sample/ standard preparation
area.
Burns to user from instrument
exhaust / chimney. Fire caused
exhaust heat to clothing /
materials.
Do not reach over the instrument for any reason.
Ensure other users are made aware when the instrument is
turned on.
Do not touch the chimney or place anything in contact with
the chimney.
Ensure the clearance room above the chimney is to that
specified in the manual.
Do not use the exhaust gases as a “heating” or “warming”
method for any material.
Do not insert your finger or any other part of body/ object/
component through the chimney inspection flap.
54