56
Residual Risk
Low (L) / Medium (M) / High (H)
Carry out regular PAT testing according to company’s PAT testing routine
and process.
Only approved persons, the manufacturer or appointed distributors are to
carry out electrical work inside the instrument.
Ensure any spills in the vicinity of the instrument or power line are cleared
up adequately and immediately and any decontamination is undertaken.
Replacing propane bottles.
Moving the instrument from storage to
area of use.
Small propane cylinders are recommended.
The instrument weighs 11kg, follow correct manual handling lifting
technique. Ask for help if required.
Refer to your companies manual handling risk assessment.
As Above “Chemical Storage”
As above “Chemical Storage”
Propane cylinder is at high pressure
Refer to the manufacturers guidelines for the safe storage, location and
operation of the gas supply cylinder.
Store and use in line with local regulations.
Hot surfaces on the flame photometer
chimney and exhaust region.
Use adequate signage to ensure other workers know the instrument is on
and hot.
See above “Burns / Fire”
Ensure you are familiar with your companies “lone working” document
and regulations. Assess as appropriate.