Chapter 3 Controls and Connectors
24 Lynx DSA User's Manual - 9240227J
• If the reset button is held for less than 7 seconds, a Factory Reset operation
will be performed. The POWER Led will blink red for few seconds, then
switch to blink green as the Lynx re-initializes and finally switch to solid
green when the Lynx is ready.
Typically a soft audible ‘click’ is heard from the front-end micro-relays as the Lynx
initializes the front-end electronics with user settings.
Following a Factory Reset operation, any initial setup
previously performed on the Lynx must be repeated because
the communication settings will have been reset to factory
state. Please refer to Lynx Setup for Administrators and
Users on page 67 for more information.