16
SECTION 8: ELECTROMAGNETIC COMPATIBILITY (EMC)
The Precise® LTM 5 Watt Soft Tissue Diode Laser needs special precautions regarding the EMC and needs to be
installed and put into service according the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect the Precise LTM. The use of accessories, transducer,
and cables other than those specied by the manufacture, CAO Group may result in increased emissions or
decreased immunity of the Precise LTM. The Precise LTM should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the Precise LTM should be observed to verify normal
operation in the conguration in which it will be used.
Guidance and manufacturer’s declaration - electromagnetic emissioins
The Precise® LTM is intended for use in the electromagnetic environment specied below.
The customer or the user of the Precise LTM laser should assure that it is used in such an environment.
Emissions Test Compliance
RF emissions
CISPR 11
Group 1
The Precise LTM laser uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not available
Voltage Fluctuations/
Flicker emissions
Not available
The Precise LTM laser is suitable for use in all establishments other than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration - electromagnetic emissioins
The Precise® LTM is intended for use in the electromagnetic environment specied below.
The customer or the user of the Precise® LTM should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short interruptions
and voltage
variations
on power supply
input lines IEC
61000-4-11
<5% UT (>95% dip in UT)
for 0,5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
<5% UT (>95% dip in UT)
for 0,5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the Precise LTM laser requires continued operation
during power mains interruptions, it is recommended
that the Precise LTM laser be powered from an
uninterruptible power supply or a battery.
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic elds should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.