18
Precise® LTM Laser Specications
Ordering Information
Weight 4.9 lbs
Dimensions
in Inches
(H x W x L)
7.6” x 8.8” x 8.75”
Laser
Classication
(Per IEC 60825)
Laser Diode; Class 4 Laser Device
Wavelength
Laser: 810 nm ± 20 nm
Aiming Beam: 630 nm - 660 nm ± 15 nm
Beam
Divergence
9 degres’s ± 1 degree
Power Range 500 mW to 5 Watts
Pulse Rate
(in Pulse Mode)
Fixed: 0.1 Hz
Pulse Mode
Duration
Fixed: 0.05 seconds
Duty Cycle
Pulse Mode: 50%
Continuous Mode: 100%
Aiming Beam
(2 mW)
YES
Audible
Notication
YES
Visual
Notication
YES
Power
Requirements
110 - 120 VAC @ 60 Hertz
220 - 240 VAC @ 50 - 60 Hertz
Amperage
1.5 Amps @ 110 - 115 VAC
0.75 Amps @ 210 - 230 VAC
Cordless
Foot Pedal
Frequency: 2.4 GHz
Battery: 9 Volt Lithium
Operating
Temperature
0 - 30°C / 32 - 86°F
Operating
& Storage
Humidity
5 - 95% RH
Accessories
Package of (20) Disposable
Handpiece Tips- Straight
002-00181
Package of (20) Disposable
Handpiece Tips- Angled
002-00182
20’ Fiber Cartridge Replacement 002-00206
Protective Glasses - Sport 002-00201
Protective Glasses - Solid 002-00202
Danger-Laser in Use Signage 002-00034
Handpiece- Autoclavable 002-00203
Fiber Stripper Tool 002-00205
Fiber Cleaver 002-00204
Indications for Uses
The Precise® LTM is approved for use by properly trained
and licensed dental professionals (as applicable) for use
in the removal of lesions, excision, incision, vaporization,
ablation, hemostasis, photocoagulation, gingivectomy,
frenectomy, operculectomy, contouring, biopsy, troughing,
ulcer care, abscess care, sulcular debridement, soft tissue
curettage, and removal of inamed edematous tissue on
soft tissue in the oral cavity.
Delivery System Specications
Quartz/Silica
Fiber
1 Cartridge - approx. 20 ft (6 meters)
Fiber Diameter 400 microns
Retractable
Fiber Delivery
Non-Autoclavable - 1
Handpiece Autoclavable - 1
Handpiece Tips Disposable: 20 (per box)
Laser Aperture YES
Warranty
Laser 2 Years Parts and Labor
Fiber and Cartridge 6 Months Parts and Labor
SECTION 9: LASER SPECIFICATIONS