4.4. GUIDED TREATMENT
In this working mode, the device establishes prefixed power output and temperature
values in function of the selected option.
Once the selected option has been activated, the device displays the treatment screen.
From that moment on, the rest of the functions are the same as for the FREE
TREATMENT mode. Please see section 4.3 FREE TREATMENT..
Indication control
Fixed values are established for the treatment time, percentage of power output and
working frequency.
There are three options (see table):
Treatment time:
10 minutes
Selected power output: 50%
Athermal
Working frequency:
0,8 MHz
Treatment time:
15 minutes
Selected power output: 75%
Moderate
thermal
Working frequency:
1,0 MHz
Treatment time:
20 minutes
Selected power output: 100%
INDICATION
CONTROL
Frankly thermal
Working frequency:
1,2 MHz
Table 4.4. Treatment parameters in case of control by indication.
Warning
Every application should be pleasant and bearable for the patient. Control
by indication does not free the operator from supervising or the patient
from giving feedback on excessive temperature increases. The patient has
the obligation to inform and to make use of the safety control