Installation and Maintenance Guide [Chapter 4]
4
25
English
Maintaining the Capsule Neuron
Capsule recommends that you carry out regular maintenance activities as described in this guide
to ensure correct operation of the device.
Periodic cleaning and disinfection
Periodically clean and disinfect the elements in the Capsule Neuron, including cables and
batteries, by wiping the outer case with a lint-free cloth, lightly moistened with warm water and a
mild, non-abrasive cleaning solution containing any of the compatible cleaning agents below
(follow the manufacturer’s guidelines):
Cleaning Precautions
Caution: When cleaning or disinfecting the Capsule Neuron, batteries, and cables, you must
power off the device first. However, you can use standard computer screen solution to
wipe down the Capsule Neuron touch screen (for example, to remove fingerprints)
during normal operation, while it is powered on.
• 70% isopropyl alcohol • 10% bleach solution
• CaviCide® • Clorox® Disinfecting Wipes
• Clorox Healthcare™ Bleach Germicidal
Cleaner
• Sani-Cloth®, Sani-Cloth® Plus, Super Sani-
Cloth®, Sani-Cloth® HB, Sani-Cloth® Bleach Wipe
• ANIOSURF Premium, Bacteranios SF,
SURFA’SAFE, WIP’ANIOS
Caution:
• To avoid damage to the Capsule Neuron, try to limit contamination and minimize the need for
disinfectants.
• Since the Capsule Neuron is not completely watertight:
- Do not spray cleaning liquid or foam on the Capsule Neuron, or submerge or soak it for cleaning.
- Never spray or squirt any type of liquid directly on the Capsule Neuron; spray the liquid, gel, or
foam on a cloth and then use that cloth to rub on the device.
- Do not spill liquids on or around the unit.
- Do not expose the power cable or power brick to liquid.
- Make sure that the cover on the USB port is securely fastened.
• Do not use any other agent than those listed. (Contact Capsule for assistance if your cleaning agent
is not on the list.)
• Do not use harsh solvents.
• Do not clean, disinfect, or sterilize any part of the system by autoclaving or with the use of ethylene
oxide gas (gas sterilization).
• Do not touch, press, or rub the display panel with abrasive cleaning compounds, instruments,
brushes or rough-surface materials.