Installation and Maintenance Guide [Chapter 5]
5
29
English
Technical Information
Guidance and manufacturers declaration
Electromagnetic Effects
The Capsule Neuron radiates and consumes radio frequency energy. It has been designed and
evaluated to comply with the emission and immunity requirements of EN/IEC 60601-1-2, but
electromagnetic interference may occur if not installed correctly and used in accordance with the
instructions in this manual. Refer to the Tables in this chapter for detailed information regarding
declaration and guidance related to emission, immunity, and separation distances.
Electromagnetic Emissions
The Capsule Neuron is intended for use in the electromagnetic environment specified below. The
customer or the user of the Capsule Neuron should make sure that it is used in such an
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Electromagnetic Immunity (ESD, EFT, Surge, Dips and Magnetic Field)
The Capsule Neuron is intended for use in the electromagnetic environment specified below. The
Caution:
Portable and mobile radio frequency devices, such as mobile phones, can impact the correct
functioning of this device and should not be used in close proximity of this device.
Use of accessories, power adapters and cables other than those specified may result in
increased emission and/or decreased immunity of the Capsule Neuron.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The Capsule Neuron uses RF energy only for its
internal function; its RF emissions are therefore very
low and are not likely to cause any interference to
nearby electronic equipment
RF emissions CISPR 11 Class A Class A: The Capsule Neuron is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Complies

Table of Contents