EasyManua.ls Logo

Carel emeter1 - Other Informations; Accuracy AV5, AV6 (in Accordance with EN50470-3 and EN62053-23); Accuracy MV5, MV6 (in Accordance with EN50470-3 and EN62053-23)

Carel emeter1
76 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ENG
24
"Carel emeter" +0300044IE rel. 1.5 - 10.11.2017
emeter3 SE
3. OTHER INFORMATIONS
3.1 Accuracy AV5, AV6 (in accordance with EN50470-3 and EN62053-23)
kWh, accuracy (RDG) according to current kvarh, accuracy (RDG) according to current
Fig. 3.a Fig. 3.b
3.2 Accuracy MV5, MV6 (in accordance with EN50470-3 and EN62053-23)
kWh, accuracy (RDG) according to current kvarh, accuracy (RDG) according to current
Limiti (Energia attiva)
Corrente di avviamento: 0,2%In
Limiti di errore di percentuale per classe index B
+1.5%
+1%
0%
-1%
-1.5%
PF=1 0.02 In 0.05 In 1.2 In
0.1 In
0.1 In
1.2 In
1.2 In
PF=L0.5
to C0.8
0.05 In
0.05 In
I
min
I
tr
I
n
I
max
Limiti (Energia Reattiva)
Co
rr
e
nt
e
d
i
a
vvi
a
m
e
nt
o
:
0,
2
%
In
Errore
sin =1 0.02 In 0.25 In 1.2A In
0.1 In 1.2A Insin =0.5
+2.5%
+2%
0%
-2%
-2.5%
0.05 In
I
min
I
tr
I
n
I
max
Fig. 3.c Fig. 3.d
Accuracy limits (active Energy)
Start-up current: 10mA
Accuracy limits (reactive Energy)
Start-up current: 0.2% In
Accuracy limits (reactive Energy)
Start-up current: 0.2% In
Accuracy limits (active Energy)
Start-up current: 10mA
Percentage limit error for index B class Error
Percentage limit error for index B class Error