35
ITA
Gas Leakage Detector +0300047IE rel. 1.1 - 26.04.2023
Ulteriori informazioni
5.2 Gas rilevati
Gruppo Registro 117 Gruppo Gas Tecnologia Dafault GAS Calibration GAS
1 R32 misti Tipo 1 Semiconduttore R32 R32
2 HFC/HFO Tipo 2 Semiconduttore R134a R134a
3 HC Tipo 3 Semiconduttore R290 R290
4 CO
2
Infrarossa CO
2
CO
2
5 NH
3
Elettrochimica NH
3
NH
3
Gas Gruppo modulo sensore Range di misura. Valore registro GasType
R-22 2 0-1000 ppm 1
R-134a 2 0-1000 ppm 2
R-404A 2 0-1000 ppm 3
R-407C 1 0-1000 ppm 4
R-410A 1 0-1000 ppm 5
R-290 3 0-4000 ppm 7
R-600A 3 0-4000 ppm 9
R-717 5 0-100 ppm 10
R-744 4 0-10000 ppm 11
R-1270 3 0-4000 ppm 13
R-407A 1 0-1000 ppm 19
R-407F 1 0-1000 ppm 22
R-32 1 0-1000 ppm 23
R-1234yf 2 0-1000 ppm 27
R-1234ze 2 0-1000 ppm 28
R-455A 1 0-1000 ppm 29
R-448A 1 0-1000 ppm 33
R-449A 1 0-1000 ppm 34
R-450A 2 0-1000 ppm 35
R-452A 1 0-1000 ppm 36
R-452B 1 0-1000 ppm 38
R-513A 2 0-1000 ppm 39
R-454B 1 0-1000 ppm 40
R-454A 1 0-1000 ppm 43
R-454C 1 0-1000 ppm 44
R-466A 1 0-1000 ppm 47
R-464A 1 0-1000 ppm 48
R-465A 1 0-1000 ppm 49
R-468A 1 0-1000 ppm 50
R-1233zde 2 0-1000 ppm 51
R-50 3 0-4000 ppm 52
R-1150 3 0-4000 ppm 53
R-507A 1 0-1000 ppm 54
5.3 Funzionamento dei relè all’accensione dello strumento
Al monento dell’accensione i 2 relè di warning e alarm si comportano nel seguente modo:
- rimangono spenti per circa 20 secondi
- si attivano per circa 2 secondi
- si disattivano per circa 2 secondi
- si riattivano se è impostata la modalità failsafe, oppure rimangono disattivati, se la modalità failsafe non è attiva.
5.4 Modalità di Funzionamento dei relè
Il funzionamento dei relè prevede 2 modalità di funzionamento, la modalità Failsafe e Modalità Warning Fault. Attraverso queste
2 modalità è possibile selezionare rispettivamente il comportamento dei relè in funzione degli allarmi previsti ed in caso di
malfunzionamento del sensore.