Detector Operation
AA5348 | 2012-12-20 2–17
Procedure to clean the battery footprint
1. Wipe the well in the detector clean of dust or debris with a soft cloth.
2. Use a soft brush or vacuum to clean out the prongs in the battery compartment well,
or contact Service for assistance.
Prongs in the Battery Compartment Well
Patient Contact
With each occurrence of patient contact:
1. Moisten a cloth with a 70 % isopropyl solution.
2. Apply the solution to the patient contact area.
Do not spray solution directly onto the equipment. Moisten a cloth with a 70 % solution of
isopropyl alcohol solution and apply to the patient contact areas after each occurrence.
Isopropyl alcohol is a flammable solvent. Read and follow instructions in the Material Safety Data
Sheet (MSDS).
If a detector is not used in a Bucky, it is enclosed in a protective plastic bag that is disposed of
after each patient exam.