Operation and Use
AF0118 | 2018-03-27 3-19
Wheels
There are two rear braked wheels and two front caster wheels.
The rear wheels are braked in normal operation. The brakes are released by
means of switches in the cart handles to allow movement. A manual release
handle releases rear wheel braking when power is not available.
X-ray Tube
Maximum Anode Heat Content = 300 kHU
Normal operation of the DRX-Revolution Nano Mobile X-ray System is considered
to be 60 exposures in a 24-hour period.
The tube has an IEC focal spot range from 0.9 mm to 1.9 mm equivalent at 0.1
mm increments.
The smaller focal spot sizes allows better image detail for smaller
patients/anatomy. Focal spot sizes are preset by the technique and are not
selectable by the user.
Other
Optionally supports DAP with automatic population.