12
n
Resolution to load cell output rate
10V impression to load cell Max.
load cell output
Recommended resolution
2 mV 1/1,000(Max)
4 mV 1/2,000(Max)
8 mV 1/5,000(Max)
Connection to AC power
Connect to the AC power and turn the power switch on.
The Input Voltage is only for 220V 50/60 Hz
External input port connection
If you are away from CI-1500A and you want to control key, Please connect the
CI-1500A with
remote key
via rear panel.
External Output port connection (realy is 5W)
Multi Connector Relay
1 ZERO RELAY
REALY OUT-PUT
2 1 STEP(LOW RELAY)
3 2 STEP(HIGH RELAY)
4 FINAL RELAY
COM RELAY OUT-PUT COM
1 ZERO KEY
KEY IN-PUT2 START KEY(F08)
COM KEY IN-PUT COM
TEST MODE
n
How to enter
Turn on the power while pressing the
ZERO
key on the front of the indicator.
4When test is done, Press
SET
key.
n
Available keys
- Set key
:
Used for moving to the next test menu.
- Other
keys
:
Used for changing the preset value.
n
Test Menu (TEST 1- TEST 6)
TEST 1 : Key test
TEST 2 : Display test
TEST 3 : Load cell test and A/D conversion test
TEST 4 : Serial interface test
TEST 5 : Printer test
TEST 6 : External input/output test (CI-1560A)
TEST 1
KEY Display Description
SET : Move to nextmenu
Other keys : Perform test
Key number
ex) In case of Zero key
1
-Press the key to be test and
the No of key mode should be
indentify with code of key as
the follows
-If you press Set key. it will be
moved to test 2.
KEY NAME No EXTERNAL INPUT No
ZERO 1 IN 1 6
TARE, 2 IN 2 7
NET/GROSS, 3
PRINT, 4
SET 5
n
FUNCTION : KEY TEST
n
KEY LIST
3
5
6