12
2.3.5. Save Setting(C–4–4)
B7 B6 B5 B4 B3 B2 B1 B0
UNIT
CHANGE
1: Possible
0: Impossible
If NTEP(C-4-1 3
rd
bit) is set, C-4-4 Setting don’t care.
2.4. Averaging A/D Value(C - 5)
(1) To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.
(2) Press the [TARE] key until the display shows “C – 5”.
(3) Press the [ZERO] key, the display shows averaging A/D.
(4) Press the [ZERO] key to finish this mode.
(5) You can change the mode by pressing [TARE] or [lb/kg(*)] key.
2.5. Real A/D Value(C – 6)
(1) To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.
(2) Press the [TARE] key until the display shows “C – 6”.
(3) Press the [ZERO] key, the display shows real A/D value.
(4) Press the [ZERO] key to finish this mode.
(5) You can change the mode by pressing [TARE] or [lb/kg(*)] key.
2.6. % Calibration (C – 7)
(1) To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.