(2) Press the [TARE] key until the display shows “C – 7”.
(3) Press the [ZERO] key, the display shows “Per 10”.
(4) Press the [lb/kg] or [*] key to have number up or down The display shows 10, 20, 30, 40, 50,
60, 70, 80 and 90 every time pressing [kg/lb] or [*]key.
If you want to calibrate PDII-150lb with 30 lb, select 20.
(5) Press the [ZERO] key, the display shows “UnLOAd”.
(6) Remove items from the platter and press the [ZERO] key. The display shows “StAbLE”
and then “LOAd”.
(7) Place 20% weight of full capacity that you setup on step (3). And press the [ZERO] key, the
display shows “StAbLE” and then “C-7End”.
(6) Remove the weight from the platter. You can change the mode by pressing [TARE] or
[lb/kg(*)] key.
2.7. Gravity Constant (C – 9)
(1) To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.
(2) Press the [TARE] key until the display shows “C – 9”.
(3) Press the [ZERO] key the display shows ”C91 “ and “ 9.7994”. Default value is 9.7994.
(4) You can change gravity constant for calibration place. Press the [lb/kg] or [*]key to have
number up or down. To move to next digit, press the [TARE] key.
(5) To save it, press the [ZERO] key. The display shows “C92” and “ 9.7994”.
(6) You can change gravity constant for using place. Press the [lb/kg] or [*] key to have number
up or down. To move to next digit, press the [TARE] key.
(7) Press the ZERO key, the display shows “C9-End” and “C – 9”.