1
CAL 3
CAL 3-1
Function: Setting Multi Calibration Step
Range of set value: 1 ~ 5
Used key Display Descriptions
STEP- 1
Setting multi calibration for step 1
(CAL3-3 and CAL 3-4 are carried out once)
STEP- 3
Setting multi calibration for step 3
(CAL3-3 and CAL 3-4 are carried out three
times.)
:Save and
next Menu navigation
: Set value change
: Changing places
STEP- 5
Setting multi calibration for step 5
(CAL3-3 and CAL 3-4 are carried out five
times.)
* If the actual curve of load cell is a straight line, set the range of weight setup to 1.
* A function to use, when the output of load cell is corrected by setting multiple points in some sections
because the actual curve of load cell is not a straight line.
(zero span)
Zero and Span points to interpolate weight from Load Cell
Multi-point
Multi-point
mV/V
Real curve
weight
(full span)
Multi-point
Ideal
Multi-point
Multi-point