9. Error Message
9-1. Error Message from the Weight Setup Mode
Error Cause Solution
Err 21
The resolution was set in excess of
the tolerance 1/30,000.
Lower the resolution.
As the resolution = maximum tolerance / value of one division,
adjust the resolution to 1/30,000 or less by correcting either the
maximum allowable weight in CAL 1 or the value of one division
in CAL3 in the weight setup mode.
Err 22
The weight for the span adjustment
was set to less than 10% of the
maximum capacity.
Set the weight to 10% or more of the maximum capacity (set in
CAL 1) from CAL 4 in the weight setup mode.
Err 23
The weight for the span adjustment
was set to more than 100% of the
maximum capacity.
Set the weight within the maximum capacity (set in CAL 1) from
CAL 4 in the weight setup mode.
Err 24
Too low span.
Set the weight again by lowering the resolution as the setting of the
current resolution is not possible because of either abnormality or
lower output in the load cell.
Two low weight for PCS and percent sample.
Err 25
Too high span.
There is either any abnormality or too high output in the load cell.
Execute steps from the zeroing step in CAL 4 in the weight set up
again.
Two high weight for PCS and percent sample.
Err 26
Too high zero point.
Check whether or not the load tray is empty.
Retry the weight setup after check at the test mode 3.
Err 27
Too low zero point.
Set the weight setting again after confirming what force is given to
the load tray of the scale in the test mode 3.
Err 28
Weight is shaking.
Check the connection of the load cell connector.