14
5.3.3 To clear indication automatically
When carrying out repeat measurement, even if it does not push the clear button after a measurement end,
a display can be cleared automatically.
1. Set up the "auto clear time" of a convenient functional numerical value. ( Reference 5.2)
2. After a measurement end, when setting time passes after a detecting element did not have load,
a display will be cleared.
* When auto clear time is set up, a display is not cleared even if it presses the "CLEAR key."
5.4 Real-time data output
A data output (real-time output mode) can be carried out about 160 times in 1 second.
It can switch every 1/ 12 seconds. The change method 6.5 Change of Output Speed of Data Reference
1Set measurement mode as "real-time output" mode.( Reference 3.1.2)
2If torque load becomes beyond the value set up beforehand,
the data output of the load torque will be carried out.
3If torque load becomes under , a data output will stop.
* The preservation to the memory of real-time data consists of 20 or more digits. When you save in memory, set
or more to 20. However, it restricts at the time of a real-time data output, and preservation of data will not be
carried out if memory becomes 800 affairs.(Overwrite is not carried out)
* During a real-time data output, since the torque display changes at high speed,
it stops being able to be visible very easily.
* When the output of data is not carried out by setup of , even if torque has started,
the torque display has become with "0."
* When is set as "0", the inside of real-time output mode continues taking out data.
The data outside the measurement range is used as a reference value.
Although the data near "0" may be fluctuated also in the time of the no-load,
it is not failure with the condition on data processing.
6 Preservation, Display, and Output of Data
Preservation, a display, and an output can be carried out for data, such as measurement data and the average.
Whenever data press the CLEAR key or require an auto clearance, as soon as memory memorizes,
they output a signal also to a data output terminal.
Statistics data output a signal also to a data output terminal at the same time they display data.
6.1 Measurement Data
6.1.1 Preservation of Measurement Data
Memory data are always saved, whenever it presses the CLEAR key or an auto clearance starts.
When a lot of data are treated, we recommend you to eliminate data before use.
* Begin the memory of data from the data number 001. When data are already memorized,