8.3 When Troubled
The checkpoint when being troubled. Before judging it as failure
When functional fault arises, check based on the following table.
By the corresponding processing, when fault is not canceled, tell repair to our company or a store.
In addition, I am allowed to make a torque tester's term of a guarantee into one year from a purchase.
(It is for a fee depending on the content of the repair.)
Use a wish attached case and a plastic Case as an object for storage of a measuring instrument.
Moreover, when you convey this product to our company by check, proofreading, repair, etc.,
for product protection, put into this case and is sure to send.
For any inquiry,
Contact your dealer.
-ISO9001-
E-mail sales@cedar.co.jp
Homepage http://www.cedar.co.jp
SUGISAKI METER CO.,LTD.
condition checkpoint solution
A power supply is not
turned on.
Is charge carried out? Carry out regular time charge using
the exclusive charger attached.
The power supply cuts
at once even if it turns
on power.
Charge is insufficient. Carry out regular time charge using
the exclusive charger attached.
It does not display, even
if it switches on a power
supply after charge of
regulation time.
If it is not used over a long
period of time, an internal
battery will carry out
full electric discharge.
An AC/DC adaptor is connected and a
system reset button is pushed.
After a return, push a clear button
and give an indication zero.
It cannot measure.
When torque is applied, does a
numerical value change?
The gauge may be damaged
when not changing. Request repair.
Is it the numerical value of
each function under setup?
End a setup and return to measurement
mode.
A display disappears
immediately.(It does not
hold by measurement
mode PP/PD)
It is used setting up an auto
clearance.
Check the setting item of an auto clear
function again. When there is no
necessity, set it as 0.0C.
A zero point cannot be
taken.
Did you clear
and make it zero?
"3.5" is seen and zero adjustment
is performed.
By real-time output
mode, even if it imposes
torque, a display does
not change.
Data are not outputted.
Torque has not reached
a real-time output start value.
At real-time output mode,
when torque is under a "CLO"
real-time output start value, a data
output does not give an indication
with "0", either. Change the value
of "CLO" into a suitable value.