Guidance and Manufacturer's declaration - Electromagnetic emissions
hyperion X9 is designed to operate in the specified electromagnetic environment. The customer or the user of
hyperion X9 must ensure its use in an electromagnetic environment with the following features:
Electromagnetic Environment
hyperion X9 uses RF energy only for its internal
operation.
For this, the RF emissions are very low and do not
interfere with the electronic devices nearby.
hyperion X9 must be used only by adequately trained
personnel (dentists and paramedics). hyperion X9
may cause radio interferences or disturb the operation
of the nearby equipment. It may be necessary to
adopt countermeasures, such as re-orienting or
moving hyperion X9 or shielding the installation site.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's declaration - Electromagnetic immunity
hyperion X9 is designed to operate in the specified electromagnetic environment. The customer or the user of
hyperion X9 must ensure its use in an electromagnetic environment with the following features:
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
Floors must be made of wood, concrete or
ceramic tiles. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
IEC 61000-4-4
fast/burst electric
transients
± 2 kV for electrical
lines
± 1 kV for input/output
lines > 3 m
The power supply line quality should be that
of a typical commercial or hospital
environment.
Overvoltage
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV standard mode
The power supply line quality should be that
of a typical commercial or hospital
environment.
Voltage drops,
short interruptions
and voltage
change on the IEC
61000-4-11 input
electric line
Ut = 0% (a 0, 45°
90°, 135°, 180°, 225°,
270°, 315°) for 0.5
cycles
Ut = 0% per 1 cycle
Ut = 70% (a 0°) per
25/30 cycles
Ut = 0% per 250/300
cycles
The power supply line quality should be that
of a typical commercial or hospital
environment. If the hyperion X9 user
requires a continuous operation also in
case of blackout, it is recommended to
power the hyperion X9 with uninterruptible
power supply or batteries.
Magnetic field at
network frequency
(50/60 Hz)
IEC 61000-4-8
The magnetic fields at network frequency
should feature levels typical of a standard
commercial or hospital environment.