OPERATOR'S MANUAL
EN
Generator reference axis:
10.3. RADIOLOGICAL FEATURES IN CBCT MODE
Operating voltage
90kV with all projections
Generator voltage at maximum current
90kV - 10mA
Load factors for the maximum output power
90kV - 10mA
Maximum output rated power (value averaged with
T>4s)
810VA (90kV 10mA - 20/25% duty cycle)
Exposure time (CBCT)
18s-36s pulsed mode with actual X-ray emission total
time 3.6s- 9s
Operating cycle in pulsed mode for CBCT exposure
20% - 90%