EasyManua.ls Logo

CEFLA NewTom VGi evo - Rotating arm maintenance; Motor test procedure

CEFLA NewTom VGi evo
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating the system
______________________________________________________________________________________
NewTom VGi evo – Service Manual rev 6 8-7
11) The analysis of the phantom scan will be
performed, showing a table with the resulting
values.
In case a value is out of range, three asterisks will
appear on its side.
Following is a short guide about what to do in case
some value is out of range.
AAP: Manually adjust the inclination of the
phantom (referring to the first scout view)
ALL: Manually adjust the inclination of the
phantom (referring to the second scout view)
RNS: Repeat blank acquisition
HDE – HDI – VDE – VDI :
1)Repeat scan
2)Repeat TIN calibration
H FWHM – V FWHM:
1)Verify phantom didn’t move
during acquisition
2)Repeat TIN calibration
HFD: 1)Repeat scan
2)Repeat TIN calibration
VFD: 1) Verify vertical alignment of the
phantom
2) Repeat TIN calibration