12 Perform a Digestion
CAUTION
If resistance is felt when trying to insert probe, ensure that the thermowell is not broken inside the Teon
cover. The sharp edges of any break can restrict the movement of the probe.
3. Load the remaining vessel assemblies into the turntable one by one. See “Vessel Distribution” section of this
manual for optimal vessel distribution.
NOTE
The sleeves should be left in the turntable at all times.
4. Ensure that each vessel is properly positioned in the turntable. The outside of the cover MUST be below the
top of the turntable.
5. Verify the oor mat positioning, ensuring that the mat is not covering the temperature sensors.