TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H • 47
7.2.1.
Temperature Settings
This section describes how to access and adjust the various temperature settings on the TEC
3000.
NOTE: Security Level 2 or higher is required to adjust the Temperature Settings.
7.2.1.1.
Enable / Disable Sensors
The default setting for both Temp A and Temp B sensors is enabled.
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
on the entry screen will make it clear which digit is being
changed. Use the “▲/▼” keys to scroll to the
appropriate number. Press “ENTER” to advance the
cursor to the next position.
2. Press “ENTER”
The display will read “Temperature Menus”.
3. Press “ENTER”
The display will read “Temp A Menu”.
NOTE: To access Temp B Menu, press “SETUP”
instead of “ENTER”.
4. Press “ENTER”
The display will read “Temp A Enabled”.
Use the “▲/▼” keys to change Temp A to DISABLED
and press “ENTER” to save the setting change.
If a setting was changed, a confirmation display will read
“New Setting Accepted”.
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Press ENTER for
Temp A Menu
or press SETUP for
next menu
Temp A
ENABLED
Use ▲▼ to adjust
Press ENTER to save
New Setting Accepted