13289499 Rev H • 58
7.2.3.3.
Lid Switch Settings
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
on the entry screen will make it clear which digit is being
changed. Use the “▲/▼” keys to scroll to the
appropriate number. Press “ENTER” to advance the
cursor to the next position.
2. Press “ENTER”
The display will read “Temperature Menus”.
3. Press “SETUP”
The display will read “Add-on Menus”.
4. Press “ENTER”
The current battery status will be displayed.
5. Press “SETUP”
Until the display reads “Lid Switch Menus”.
6. Press “ENTER”
The display will read “Lid Switch Installed”. Use the
“▲/▼” keys to adjust and press “ENTER” to save.
7. Press “SETUP”
The display will read “Lid Open Fog Clear”. Use the
“▲/▼” keys to adjust and press “ENTER” to save.
8. Press “SETUP”
The display will read “Lid Open Alarm Time”. Use the
“▲/▼” keys to adjust the value and press “ENTER” to
save.
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Press ENTER for
Add-on menus
or press SETUP for
next menu
Battery Status
On AC Power
Press ENTER for
Lid Switch menus
or press SETUP for
next menu
Lid Switch Installed
NO
Use ▲▼ to adjust
Press ENTER to save
Lid Open Fog Clear
DISABLED
Use ▲▼ to adjust
Press ENTER to save
Lid Open Alarm Time
1 min
Use ▲▼ to adjust
Press ENTER to save