MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H • 62
7.2.4.4. Languages
The TEC 3000 has five language options to choose from: English, Spanish, German,
Italian, and French.
NOTE: Security Level 2 or higher is required to change the language setting.
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
on the entry screen will make it clear which digit is being
changed. Use the “▲/▼” keys to scroll to the
appropriate number. Press “ENTER” to advance the
cursor to the next position.
2. Press “ENTER”
The display will read “Temperature Menus”.
3. Press “SETUP”
Until the display reads “Display and Output”.
4. Press “ENTER”
The display will read “Temperature Units”.
5. Press “SETUP”
Until the display reads “Advanced Display and Output”.
6. Press “ENTER”
The display will read “Alarm Buzzer”.
7. Press “SETUP”
The display will read “Language”. Use the “▲/▼” keys to
adjust and press “ENTER” to save.
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Press ENTER for
Display and Output
or press SETUP for
next menu
Temperature Units
°C
Use ▲▼ to adjust
Press ENTER to save
Press ENTER for
Advanced Display
and Output or press
SETUP for next menu
Alarm Buzzer
ENABLED
Use ▲▼ to adjust
Press ENTER to save
Language
ENGLISH
Use ▲▼ to adjust
Press ENTER to save