MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H • 50
7. Press “SETUP”
The display will read “Temp A High Alarm”.
Use the “▲/▼” keys to adjust setting and press
“ENTER” to save the change.
8. Press “SETUP”
The display will read “Temp A Low Alarm”.
Use the “▲/▼” keys to adjust setting and press
“ENTER” to save the change.
Temp A High Alarm
ENABLED
Use ▲▼ to adjust
Press ENTER to save
Temp A Low Alarm
ENABLED
Use ▲▼ to adjust
Press ENTER to save