MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H • 65
5. Press “ENTER”
The display will read “Timed Fill”. Use the “▲/▼” keys to
adjust and press “ENTER” to save.
6. Press “SETUP”
The display will read “Timed Fill Frequency”. Use the
“▲/▼” keys to adjust and press “ENTER” to save. It can
be set from 1 to 28 days.
7. Press “SETUP”
The display will read “Timed Fill Start Hour”. Use the
“▲/▼” keys to adjust and press “ENTER” to save.
8. Press “SETUP”
The display will read “Timed Fill Start Minute” Use the
“▲/▼” keys to adjust and press “ENTER” to save.
Timed Fill
DISABLED
Use ▲▼ to adjust
Press ENTER to save
Timed Fill Frequency
5 Days
Use ▲▼ to adjust
Press ENTER to save
Timed Fill Start Hour
20:00
Use ▲▼ to adjust
Press ENTER to save
Timed Fill Start Minute
20:00
Use ▲▼ to adjust
Press ENTER to save