troubleshooting should not be necessary. If problems do
If you are not familiar with normal Trifecta XPro Series
before continuing.
things that should be monitored or checked for inconsistency
to aid in diagnosis:
1.
Series are in their normal operating position.
2.
the instructions in the Operations section of the manual.
3. Check for frost formation on lines which should not
have recently been active.
4. Leak check all plumbing components.
5. Check for discrepancies between analog gauges and
controller parameters.
The above steps are described in greater detail below and
will lead you to the root of the malfunction.
Conrm Valve Positions
•
are in the correct position:
– Bulk Tank Auxiliary Liquid Valve Open
– Bulk Tank Pressure Building Valve Open
•
Series
• Manual override hand wheel in the “Out” position
Normal Activation of Solenoids
Dial main switch on front of the control panel to “Monitor.”
Open control panel door giving you access to the PLC
each cluster of solenoids. Make sure all valves energize
accordingly.
Troubleshooting
Frost Formation
through them. The frost will begin to thaw when product is
can often be detected.
Leak Check
to plumbing stack on tank and all connections to pressure
transmitters.
• A leak at any point in the Trifecta XPro Series system
will result in loss of product.
• A leak in the plumbing within the plumbing stack or
pressure transducers and transmitters can cause incorrect
values to be translated by the PLC. These transducers
which tell it when to open and close solenoids. Incorrect
values will cause the Trifecta unit to operate in a random
fashion and could cause supply issues downstream.
Data from Main Screen and Analog
Pressure Gauges
Record Trifecta XPro Series and bulk tank parameters as
described in the Parameter Settings section of this manual.
Compare these values to those on the analog gauges of the
tanks and the bulk tank. (an example form is on the next
page.)