DIRECTIONS FOR USE PATIENT
Language: To switch the language of the device:
. Select “CONFIG” on the Home screen by pressing the
lower right-hand soft key
. Select “SETTINGS” by pressing the upper right-hand
soft key
. Press the lower left-hand soft key once to toggle to the
“Language” option
. Use the right-hand soft keys to select your preferred
language
. Press “OK” on the bottom right
Contrast: To adjust the contrast of the screen:
. Select “CONFIG” on the Home screen by pressing the
lower right-hand soft key
. Select “SETTINGS” by pressing the upper right-hand soft
key
. The screen will default to the “Contrast” option
. Use the right-hand soft keys to adjust the contrast as
desired
. Press “OK” on the bottom right
TROUBLESHOOTING AND REPAIR
Troubleshooting
If the device does not function:
Make sure the batteries are properly installed
(check polarity markings)
Make sure the battery contacts are clean.
If the device is ON, but does not respond to pressing the
key pad buttons:
Detach all patient lead wires from the device.
Remove batteries from the device.
Wait  seconds.
Re-insert batteries and resume treatment.
If the Low Battery Indicator is visible, replace both
batteries.
If the device is on, the indicator lights are illuminated over
the intensity controls, and you feel no stimulation, check
and verify the connection of lead wires and electrodes.
If the device appears to be functioning, and there is no
stimulation, replace the lead wires and/or electrodes.
If the channel shows an “OPEN” error message, check to
make sure that all electrodes and lead wires are properly
attached.
If the message continues to show after all electrodes and
wires have been checked, remove the electrodes, clean
and dry the skin area under the electrodes, and replace the
electrodes.
If the error message persists, call Chattanooga Customer
Service.
If you are unable to increase the intensity of the device, the
device’s intensity lock safety feature has likely engaged.
To disengage, decrease the channel’s intensity. You should
then be able to increase the intensity.
For more detailed Troubleshooting information, see
“Troubleshooting” on page .
Repair
There are no user serviceable parts inside the device.
If the device appears to be non-functional, contact your
clinician, or contact Chattanooga directly
MAINTENANCE, CLEANING, AND STORAGE
Maintenance
Under normal conditions, the device does not require
periodic maintenance, calibration or testing.
Cleaning
Use a damp cloth with mild soap to clean the exterior of
the device and lead wires. Use of other cleaning solutions
may damage these items. Never immerse the device in
liquids.
Storage
To store the stimulator for an extended time (more than
 days), remove the batteries and store the device in a
cool, dry place.