24
PRODUCT INFORMATION CLINICIAN
Changing the Switch Working Mode
The Chattanooga Continuum has the ability to modify
the working mode of an attached Hand or Foot switch.
Available options are “Simultaneous” or “Alternated”.
In “Simutaneous” mode, both channel  and 
will work simultaneously.
In “Alternated” mode, channel  and  will work
alternately
To access this setting, the following steps are to be applied:
. Press the ON button to turn on the device.
. Once the main Options menu is displayed, choose
“CONFIG” option.
. Choose the “SETTINGS” option.
. Use the lower left soft key to scroll to the “SWITCH”
option.
. Use the right side soft keys to select the switch working
mode
. Press the home screen button.
. Notice that if a switch is plugged in, an icon will be
shown in the upper right corner of the display.
Using the Foot Switch
The Chattanooga Continuum unit has the ability to use
a foot switch for gait training. The Foot Switch must
be plugged in so that the unit recognizes that a gait
treatment is wanted. Otherwise, a normal Large Muscle,
Custom treatment is performed. To do this, the following
steps are to be applied:
. Press the ON button to turn on the device.
. Once the main Options menu is displayed, choose the
NMES option.
. Choose the “LARGE MUSCLE” option.
. Choose the “CUSTOM” option.
. Modify adjustable regimen parameters as needed and
then press the “OK” button.
. Plug the Foot Switch into the remote jack in the unit.
. Increase intensity. For Channel , the unit will output a
waveform as the foot is lifted up, and stop its output as
the foot is down. For Channel , the output will behave
according to the selected switch working mode.
Both switches should be used with the custom programs,
not with the preset therapies.
ACCESSORIES
Electrodes
Any Chattanooga Electrode with a minimum active area
of  cm may be used with this device. This includes
Chattanooga square cm x cm (” x ”) StimCare
electrodes.
DANGER: Use of an electrode with an area less than 
cm can cause burns when the unit is used at higher
intensities. Consult your clinician prior to using any
electrode less than  cm.
Lead Wires
Any Chattanooga TENS/NMES Lead Wire ( cm - ”)
with the touch proof device connection (part number
).
Remote Switches
Use either: Chattanooga Foot Switch, Available
as an optional accessory to the Continuum
device - Part Number: 
Chattanooga Hand Switch supplied as standard
with Continuum device - Part Number: 