TENS Back — Summary
This treatment option is a -channel Modulated Amplitude
program for the management of acute or chronic pain
control in the Back. All parameters are preset with the
exception of Treatment time and the Pulse Rate, which
can be adjusted by the clinician.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Pulse Rate 2 Hz-150 Hz 100 Hz
Preset Parameters Settings
Cycling Time 12 sec
Attenuated Span 40%
Pulse Width 300 μs
Mode MOD
TENS Hand — Summary
This treatment option is a -channel Simple Modulated
Pulse (SMP) program for the management of acute or
chronic pain control in the hand. This treatment option
modulates the pulse rate and the pulse width at the
same time. All parameters are preset with the exception
of Treatment Time and the Pulse Rate , which can be
adjusted by the clinician.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Pulse Rate 2 Hz-150 Hz 100 Hz
Preset Parameters Settings
Cycling Time 12 sec
Pulse Width 300 μs
Mode SMP
TENS Foot — Summary
This treatment option is a -channel Simple Modulated
Pulse (SMP) program for the management of acute or
chronic pain control in the foot. This treatment option
modulates the pulse rate and the pulse width at the
same time. All parameters are preset with the exception
of Treatment Time and the Pulse Rate , which can be
adjusted by the clinician.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Pulse Rate 2 Hz-150 Hz 100 Hz
Preset Parameters Settings
Cycling Time 12 sec
Pulse Width 300 μs
Mode SMP
TENS Custom — Summary
This treatment option is a sequenced, -channel
Modulated Amplitude (MOD) or a Simple Modulated Pulse
(SMP) program for the management of acute or chronic
pain.
All parameters have default settings and are adjustable by
the clinician within the range described in the table below.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Pulse Rate 2 Hz-150 Hz 100 Hz
Cycling Time 3-20 sec 12 sec
Span (MOD Only) 20% - 95% 40%
Mode SMP, MOD SMP
Preset Parameters Settings
Pulse Width 300 μs