FINAL TEST
EN
| Intelect
®
Mobile 2 - Service Manual 52
8.1.1. Tracker sheet for Intelect Mobile 2 ULTRASOUND Final Test Procedure
1. Serial Number (S/N) Done
2. Software Version: ⎵⎵.⎵⎵.⎵⎵ Done
Ultrasound
3. Ultrasound therapy runs without error. Pass Fail
4. The display indicates “Applicator disconnected (0x00000000)”. Pass Fail
5. The ultrasound program is resumed. Pass Fail
Leakage Current
6. The Voltage DC is between 0mV and 10mV after stabilization. Pass Fail
The Voltage AC is between 1mV and 100 mV after stabilization. Pass Fail
7. The Voltage DC is between 0mV and 10mV after stabilization. Pass Fail
The Voltage AC is between 1mV and 100 mV after stabilization. Pass Fail
Battery (test optional)
8. The battery is detected. Pass Fail
Visual Check
9. All tests performed are passed and the battery is removed. Pass Fail
Current date (YYYY MM DD): / /
Operator name: _______________________________________
Operator signature: ____________________________________