FINAL TEST
EN
| Intelect
®
Mobile 2 - Service Manual 54
Press on intensity button and increase the current
with the knob for two complete rounds (> 5mA),
then press the “Play” button.
You should have the CH2 status channel
“Running”.
Check if the both LED around the connectors are
ON and blinking.
Let the programs “Running” running for 15
seconds and verify that there is not pause or any
popped error.
Check Pass/Fail the tracker sheet §3 if the program
runs on both channels without problem.
Simulation of Electrode fault on CH2
Remove the Electrotherapy connector 2 from the
device, an advice is displayed indicating “Electrode
fault (0x71020000)”.
Check Pass/Fail the tracker sheet §4 if “Electrode
fault (0x71020000)” is displayed.
Resumption of the Electrotherapy program on
CH2
Put back the Electrotherapy connector 2 into the
device.
Press Resume.
You should have the CH2 status channel
“Running”.
Press “Intensity” field and increase the value with
the knob until the maximum, 100 mA.
Check Pass/Fail the tracker sheet §5 if the program
is resumed without error.