EasyManua.ls Logo

Check-line DTS-200 - Page 18

Check-line DTS-200
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
–17 –
Step 2:
Set the diameter
This step is only required for tension meters with a material
thickness compensator
Step 3:
Perform a zero adjustment with the tension meter in its measuring
position
Step 4:
Calibrate calibration point 1
Insert material to be measured
Step 5:
Calibrate calibration point 2
Step 6:
Calibrate calibration point 3
Step 7:
Exit the calibration process or calibrate a second diameter. In this
case, continue with step 2.
NOTE: By calibrating a second diameter, you can dene a material thickness range to be used
for the material, e.g. 0.1 – 0.3 mm. Therefore the tension meter can be used for a diameter
range instead of a specic diameter only.
Once you have completed the calibration, make sure to verify it as described in
Section 6.2. In case of a large deviation, repeat the calibration or select different
calibration points.

Related product manuals