– 8 –
Material Setup Submenu Settings Menu Description
No. of the
characteristic-
material curve
1 - 5
[character],
[numbers],
special character]
To enter a name for the selected material
characteristic.
Damping [1] – [9]; [5] Section 5.5
Alarms
[ON], [OFF] To activate or deactivate the alarm function for the
material characteristic.
High Limit
[0000] - [9999] Activate/deactivate the alarm for all calibrations.
The alarm of a material characteristic is only active if
the [Alarm] menu item in Material Setup is activated
as well.
Low Limit [0000] - [9999] Activate/deactivate the peak values in the bar graph
display.
Calibration 6.1
Start
To perform a calibration, follow the instructions shown
on the display.
Cal. Points
between [10 %] and
[100 %] of the tension
range
Set three calibration points for which a calibration
should be performed.
Weights [cN/daN] [g/kg] Set the unit of the calibration weights used.
4.3 Material Menu
In the material menu you can make the settings for the selected material characteristics
and perform the calibration. To perform the calibration, the weights for the selected
calibration points must be available.
4.4 Password
Setting a password blocks the access to some menu parts for unauthorized users:
0000 All menus are accessible.
0001 – 0999 The Material, Cal. Adjustment, Settings, and Memory Settings menus
are locked
1000 – 9999 The entire menu is locked.
NOTE: To apply a changed password, exit the main menu. Entering 0000 as the
password disables the password feature. If you forgot the password, please contact
the factory to request the master password. As an alternative, you can perform a factory
reset. Please note, that in this case, you will lose any customer-dened calibrations.
4.5 Factory Reset
A factory reset resets the tension meter to its original manufacturer settings.
This procedure will delete all settings, including any customer-dened material
characteristics (calibrations); the factory calibration, however, will be kept.
Customer calibrations will be deleted.