CI-154 User’s Manual
Page 51
x=1 to enable or 0 to disable confirmation screen with
message:
“CONFIGURATION OF COUNTER WILL BE CHANGED”
[CONFIRM] [CANCEL]
If enabled, BAT command execution will require user
confirmation. ACK1 argument must be the first command in
the BAT series of commands. Setting of ACK1 is persistent
but reverts to 0 upon power up.
Pressing [CONFIRM] executes BAT command. Pressing
[CANCEL] cancels execution of the BAT command.
Set sample volume in cubic feet (CF), range B0.1F to B200F.
Set sample volume in cubic meter (CM), range B0.01M to
20CM.
Set sample volume in liter (L), range B5L to B9999L.
Execute batch of commands separated by | (7C hex).
Maximum length of command string is 99 characters.
Example:
BAT|ACT1=1|B50L|I15|P0|G3|Karea 1|$d0101|$p0101|S
Enable confirmation screen and if CONFIRM is pressed:
Set sample volume to 50 liters.
Set initial delay to 15 seconds.
Set delay between samples to 0.
Set number of samples to 3
Set ID name to Area 1
Display first (smallest) and third channels
Print first and third channels.
Start sampling.
Set initial delay, range I02 to I23:59:59.
(Example: I01:00 sets initial delay to 1 minute.)
Set delay between samples, range P0 to P23:59:59.
(Example: P00:00:15 or P00:15 sets delay to 15 seconds.) If
nnnn < or = sample time, the delay is end to start; if nnnn
> than the sample time, the delay is start to start.
Set number of samples. Range is 1 to 254. (Also sets to stop
after n. Sets number of samples to infinite if length of nnn
string is 0.)
Set date/time in 2 digit month/day/year/hour/minute/second
format.
Clear entire ID table. This must be followed immediately by
a <CR>.
Set current ID to 16-character string. Insert string as a 16-
character ID into the ID table if it does not already exist.
String length may vary from one to 16 characters. Valid
characters are the same as those available from the front
panel interface.
Set/Add Current ID, up to 16 character string. Does NOT add
ID to program memory or blank out program name.