CI-154 User’s Manual
Page 61
To read the location ID registers of the most recent sample, issue the command as
follows:
MBAP HEADER
FUNCTION CODE
DATA
Sent to unit
00 01 00 00 00 06 01
03
00 64 00 08
Received from unit
00 01 00 00 00 13 01
03
10
65 64 61 66 6c 75 20 74 64
69 20 20 20 20 20 20
Note: the 1
st
byte in the DATA field is a hex 10 to indicate there are 16 bytes that
follow.