Cole Palmer 4-7
Section 4 Operation
Settings - Application Settings is used to view/adjust the controller's ve Setpoints and Real
Temperature Adjustments (RTA) enable/disable the alarms, change the uid type, set the pump speed,
congure the interfaces (optional), set the clock, turn the timer on or off, and turn auto restart and en-
ergy savings on or off.
1. With Application Settings highlighted
press to view:
Menu
Setpoints
Alarms
FluidType
PumpAdjustment
Menu
Interfaces
Set Clock
On/OffTimer
AutoRestart
Scroll down for additional options:
2. With Setpoints highlighted, press
to display the list. Use the up/down arrows to
highlight the desired SP. NOTE Use the down
arrow to display SP5.
3. Press .
Menu
SP1 xx.x
RTA int xx.x
Menu
SP1
SP2
SP3
SP4
If this temperature on the Start/Status Displaysdoes not accurately reect the actual temperature in the
reservoir, an RTA can be applied. The RTA can be set ±10°C (±18°F).
As an example, if the unit's temperature is stabilized and displaying 20°C but a calibrated reference
thermometer reads 20.5°C, the RTA should be set to -0.5°C. After you enter a RTA value allow unit to
stabilize before verifying the temperature in the bath. NOTE If display accuracy is required, we recom-
mend repeating this procedure at various setpoint temperatures and on a regular basis.
NOTE You cannot adjust the setpoint closer than 0.1°C to either of the uid's system limits, see Fluids
Type in this Section.
4. With the desired line highlighted press .
The right-most digit will have a cursor beneath it.
Use the left and right arrows to move the cursor
to the desired digit and then use the up and down
arrows to change the value. Once all the desired
changes are made, press to save the change.
or to cancel it.
35.0°C